Hardware and/or software problem – Watchdog timer reset error.
1. Poor electrical connection or faulty wiring (bad power source to
sterilizer).
2. Main PC board may be faulty.
Software interruption error.
Run additional cycles. If problem continues, replace the Main PC
board.
Ram Test Error.
Run additional cycles. If problem continues, replace the Main PC
board.
Door malfunction: Error Code 8 Door switch did not return to normally
open condition after door pulse solenoid actuated. Error Code 9 –
Display Pressure exceeded 0.9 PSI during the Dry cycle.
If pressure reading is not at zero (0), sterilizer needs to be re-calibrated.
Refer to 5.0 - Calibration Procedure.
If door opens at end of cycle, check door switch continuity (normally
open).
If door does not open at end of cycle, refer to 4.0 - Door Opening Test
Procedure.
Inspect door mechanism for smooth operation. Lubricate if necessary.
Check door switch and wire connections.
Check door pulse solenoid.
Pressure display reads less than 24.5 psi.
Check for pressure leaks. Refer to 1.0 - Leak Detection Guide.
Temperature probe is dirty or malfunctioning.
Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
ERROR CODE 7 ERROR CODE 10
ERROR CODE 12
ERROR CODE 11
ERROR CODE 8 or 9
3.0 - ERROR CODES AND OTHER DIAGNOSTICS (continued)