4.0 - DOOR OPENING TEST PROCEDURE
Note: This test will only verify that the door opening relay on the main PC board is functioning. A complete
cycle is necessary to verify function of the sensors and the latching mechanism.
1. To execute test, start any sterilization cycle. After chamber has filled with water, the "sterilizing" LED will light up.
2. Disconnect the power cord.
3. Open the door.
4. Depress STOP and POUCHES button simultaneously while re-connecting the power cord.
5. Sequentially depress each button from left to right across the panel.
6. Close the door. In sequence, this will make active all relays on the Main PC board.
7. To repeat this procedure, simply repeat Steps #2 through #6.