An ISO 9001 Certified CompanyRPI FIELD SERVICECALIBRATION KITRPI Part #MIK074For servicing the Midmark • Ritter Ultra Clave®CALL (800) 221-9723 or (81
9AT-A-GLANCECALIBRATION PROCEDURE1) Open door.2) Depress LIQUIDSand PACKS buttons at the same time as plugging in thepower cord.3.) Starting with UNWR
10Figure 1 – Diagnostics Procedure (M9 & M11 and M9D & M11D Style Key Pad Displays)M9 & M11 Style Key Pad DisplayM9D & M11D Style Key
5.0 - CALIBRATION PROCEDURE - PRESSURE/TEMPERATURE POTENTIOMETER ADJUSTMENTS (continued)11NOTES: 1) Turning the adjusting screw of the pressure range
8) CONNECTING THE TEST PRESSURE GAUGE TO THE STERILIZER: Disconnect the black tubing from Pressure Sensor and from the barb fitting.Keep black tubing
13Figure 2 – Old Style PC BoardNOTE: The pressure zero, pressure range, and temperature potentiometers must not be adjusted exclusively. If one potent
14NOTE: The pressure zero, pressure range, and temperature potentiometers must not be adjusted exclusively. If one potentiometer needs to be adjusted,
RPI PART # OEM PART # DESCRIPTIONMIP050 002-0357-00 TEMPERATURE PROBE ASSEMBLY • •MIS075 002-0358-00 WATER LEVEL SENSOR ASSY • •PCV052 002-0359-00 PRE
This guide has been prepared to help troubleshoot, evaluate and repair the Midmark•Ritter M9 and M11 Ultra-Clave®.TABLE OF CONTENTS1.0 – Leak Detectio
21.0 - LEAK DETECTION GUIDEDOOR and/or DAM GASKET Check for water or steam leaking around the door.VENT SOLENOID VALVE Check for water or steam leakin
FILTERS: MIF038 or MIF062Depending on the age of the machine, you will need to replaceeither the sintered bronze style Chamber Filter (MIF038) or both
4Power interruption to control PC board.1. Overheat thermostat activates (opens).• Pressure leak during sterilization cycle.• Water sensor shorted (wh
5Hardware and/or software problem – Watchdog timer reset error.1. Poor electrical connection or faulty wiring (bad power source tosterilizer).2. Main
6Printer FuseFast Acting; 1/4" dia. x 1-1/4" Lg.; 3 Amp, 250 VoltLine Fuse(s)Fast Acting; 1/4" dia. x 1-1/4" Lg.; 15 Amp, 250 Volt
74.0 - DOOR OPENING TEST PROCEDURENote: This test will only verify that the door opening relay on the main PC board is functioning. A completecycle is
8IMPORTANT The Calibration Procedure that begins onpage 9 requires the sterilizer to be plugged in andpowered up while the adjustments are being per-